Recall Z-1931-2015— Roche Diagnostics Operations, Inc.
Class II · Terminated
Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2015-05-07, terminated 2017-03-24. It names one FDA 510(k) clearance: K873363. Product: Uric Acid plus ver.2 In vitro test for the quantitative determination of uric acid in human serum, plasma and urine or Roche/Hitachi MODULAR systems, cobas c111, COBAS Integra 800 and 400 plus systems, and cobas c systems.. Reason: Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine..
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-05-07
- Terminated
- 2017-03-24
- Firm location
- Indianapolis, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Uric Acid plus ver.2 In vitro test for the quantitative determination of uric acid in human serum, plasma and urine or Roche/Hitachi MODULAR systems, cobas c111, COBAS Integra 800 and 400 plus systems, and cobas c systems.
Reason for recall
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
Data sourced from openFDA. This site is unofficial and independent of the FDA.