Recall Z-1930-2015— Roche Diagnostics Operations, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2015-05-07, terminated 2017-03-24. It names one FDA 510(k) clearance: K873049. Product: Triglycerides/Glycerol Blanked. In vitro test for the quantitative determination of triglycerides in human serum and plasma or Roche/Hitachi MODULAR systems, and cobas c systems.. Reason: Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine..

Recalling firm
Roche Diagnostics Operations, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-05-07
Terminated
2017-03-24
Firm location
Indianapolis, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Triglycerides/Glycerol Blanked. In vitro test for the quantitative determination of triglycerides in human serum and plasma or Roche/Hitachi MODULAR systems, and cobas c systems.

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Data sourced from openFDA. This site is unofficial and independent of the FDA.