Recall Z-1929-2025— Philips North America

Class II · Ongoing

Class II FDA medical device recall by Philips North America, initiated 2025-05-29. It names one FDA 510(k) clearance: K193215. Product: Intera 1.5T Achieva IT Nova Product Number: 781175. Reason: identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage..

Recalling firm
Philips North America
Classification
Class II
Status
Ongoing
Initiated
2025-05-29
Firm location
Cambridge, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Intera 1.5T Achieva IT Nova Product Number: 781175

Reason for recall

identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.

Data sourced from openFDA. This site is unofficial and independent of the FDA.