Recall Z-1928-2014— Alere San Diego, Inc.
Class III · Terminated
Class III FDA medical device recall by Alere San Diego, Inc., initiated 2014-05-30, terminated 2015-02-11. FDA's recall record names no specific 510(k) clearance for this event. Product: Alere Triage BNP Calibrators for the Beckman Coulter Access Family of Immunoassay Systems, Model No. 98202 The Alere Triage BNP test is intended for use with the Beckman Coulter Access Family of Immunoassay Systems for the In Vitro quantitative measurement of B-type natriuretic peptide (BNP) in plasma specimens using EDTA as the anticoagulant. The test is intended to be used for the following indications: ¿ as an aid in the diagnosis of congestive heart failure (also referred to as heart failure) ¿ as an aid in the assessment of severity of congestive heart failure ¿ for the risk stratification of patients with acute coronary syndromes ¿ for the risk stratification of patients with heart failure The Alere Triage BNP Calibrators are intended to calibrate the Alere Triage BNP test for the quantitative determination of BNP levels in human EDTA plasma using the Beckman Coulter Access Family of Immunoassay Systems.. Reason: Alere San Diego is recalling Alere Triage BNP Calibrators for the Beckman Coulter Access Family of Immunoassay Systems because the bottle labeled as Calibrator Level 5 actually contains Calibrator Level 4. This results in a calibration failure when attempting to calibrate using a calibrator kit which contains the affected Level 5 Calibrator..
- Recalling firm
- Alere San Diego, Inc.
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2014-05-30
- Terminated
- 2015-02-11
- Firm location
- San Diego, CA, United States
Product description
Alere Triage BNP Calibrators for the Beckman Coulter Access Family of Immunoassay Systems, Model No. 98202 The Alere Triage BNP test is intended for use with the Beckman Coulter Access Family of Immunoassay Systems for the In Vitro quantitative measurement of B-type natriuretic peptide (BNP) in plasma specimens using EDTA as the anticoagulant. The test is intended to be used for the following indications: ¿ as an aid in the diagnosis of congestive heart failure (also referred to as heart failure) ¿ as an aid in the assessment of severity of congestive heart failure ¿ for the risk stratification of patients with acute coronary syndromes ¿ for the risk stratification of patients with heart failure The Alere Triage BNP Calibrators are intended to calibrate the Alere Triage BNP test for the quantitative determination of BNP levels in human EDTA plasma using the Beckman Coulter Access Family of Immunoassay Systems.
Reason for recall
Alere San Diego is recalling Alere Triage BNP Calibrators for the Beckman Coulter Access Family of Immunoassay Systems because the bottle labeled as Calibrator Level 5 actually contains Calibrator Level 4. This results in a calibration failure when attempting to calibrate using a calibrator kit which contains the affected Level 5 Calibrator.
Data sourced from openFDA. This site is unofficial and independent of the FDA.