Recall Z-1927-2014— Siemens Healthcare Diagnostics, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc, initiated 2014-06-02, terminated 2015-02-05. FDA's recall record names no specific 510(k) clearance for this event. Product: Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot number: 454778 Kit Catalog Number: 112219 SMN: 10310026 Intended Use: In vitro diagnostic use. Acid Regent is a component of the ADVIA Centaur and ADVIA Centaur XP instruments. Reason: Swollen acid bottles due to a trace metal contamination in the Acid Reagent bottles.

Recalling firm
Siemens Healthcare Diagnostics, Inc
Classification
Class II
Status
Terminated
Initiated
2014-06-02
Terminated
2015-02-05
Firm location
East Walpole, MA, United States

Product description

Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot number: 454778 Kit Catalog Number: 112219 SMN: 10310026 Intended Use: In vitro diagnostic use. Acid Regent is a component of the ADVIA Centaur and ADVIA Centaur XP instruments

Reason for recall

Swollen acid bottles due to a trace metal contamination in the Acid Reagent bottles

Data sourced from openFDA. This site is unofficial and independent of the FDA.