Recall Z-1924-2013— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2013-07-01, terminated 2014-08-26. It names one FDA 510(k) clearance: K060502. Product: Dimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (DC107) Product Usage: The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of tacrolimus therapy in kidney and liver transplant patients.. Reason: Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimension TACR Flex Reagent cartridge (DF107) lot BB4087 and its linked calibrator (DC 107) lot 3BD029. Falsely low Tacrolimus results across the assay range may lead to alterations in dosing..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-07-01
Terminated
2014-08-26
Firm location
Newark, DE, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Dimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (DC107) Product Usage: The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of tacrolimus therapy in kidney and liver transplant patients.

Reason for recall

Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimension TACR Flex Reagent cartridge (DF107) lot BB4087 and its linked calibrator (DC 107) lot 3BD029. Falsely low Tacrolimus results across the assay range may lead to alterations in dosing.

Data sourced from openFDA. This site is unofficial and independent of the FDA.