Recall Z-1923-2019— Philips North America, LLC
Class II · Terminated
Class II FDA medical device recall by Philips North America, LLC, initiated 2019-06-07, terminated 2020-05-08. It names one FDA 510(k) clearance: K063375. Product: Philips HeartStart MRx devices in use with the M3539A AC Power Module: M3535A, M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, M3536M9. Reason: The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring and failure or delay in the delivery of a shock or pacing can occur. A faulty AC Power Module may also fail to charge the battery..
- Recalling firm
- Philips North America, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-06-07
- Terminated
- 2020-05-08
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips HeartStart MRx devices in use with the M3539A AC Power Module: M3535A, M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, M3536M9
Reason for recall
The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring and failure or delay in the delivery of a shock or pacing can occur. A faulty AC Power Module may also fail to charge the battery.