Recall Z-1921-2018— Smith & Nephew, Inc.

Class II · Terminated

Class II FDA medical device recall by Smith & Nephew, Inc., initiated 2018-05-02, terminated 2020-09-01. It names one FDA 510(k) clearance: K052792. Product: smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.. Reason: Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm..

Recalling firm
Smith & Nephew, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-05-02
Terminated
2020-09-01
Firm location
Memphis, TN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.

Reason for recall

Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.