Recall Z-1920-2020— ICU Medical, Inc.
Class II · Terminated
Class II FDA medical device recall by ICU Medical, Inc., initiated 2019-10-04, terminated 2021-12-28. It names one FDA 510(k) clearance: K964435. Product: REF: Z3806, 131" (333 cm) 10 Drop Primary Set w/2 MicroClave, Remv 4 Gang 4-way Stopcocks (1 Remv), Check Valve, Rotating Luer, 1 Ext, Sterile R, UDI: (01)00887709059108. Reason: There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences..
- Recalling firm
- ICU Medical, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-10-04
- Terminated
- 2021-12-28
- Firm location
- San Clemente, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
REF: Z3806, 131" (333 cm) 10 Drop Primary Set w/2 MicroClave, Remv 4 Gang 4-way Stopcocks (1 Remv), Check Valve, Rotating Luer, 1 Ext, Sterile R, UDI: (01)00887709059108
Reason for recall
There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.