Recall Z-1917-2020— ICU Medical, Inc.

Class II · Terminated

Class II FDA medical device recall by ICU Medical, Inc., initiated 2019-10-04, terminated 2021-12-28. It names one FDA 510(k) clearance: K964435. Product: REF: Z2812, 121" 10 Drop Primary Set w/5 MicroClave, Remv 2 Gang 4-Way Stopcocks, Rotating Luer, 2 Ext, Sterile R, UDI: (01)00887709028074. Reason: There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences..

Recalling firm
ICU Medical, Inc.
Classification
Class II
Status
Terminated
Initiated
2019-10-04
Terminated
2021-12-28
Firm location
San Clemente, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

REF: Z2812, 121" 10 Drop Primary Set w/5 MicroClave, Remv 2 Gang 4-Way Stopcocks, Rotating Luer, 2 Ext, Sterile R, UDI: (01)00887709028074

Reason for recall

There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.

Data sourced from openFDA. This site is unofficial and independent of the FDA.