Recall Z-1917-2013— Biomerieux Inc
Class II · Terminated
Class II FDA medical device recall by Biomerieux Inc, initiated 2013-06-14, terminated 2015-06-01. FDA's recall record names no specific 510(k) clearance for this event. Product: bioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDT V2S 6.01 PT. Part number: 414566. The product is used with the VITEK 2 automated system which consists of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.. Reason: A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bid-directional Computer interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin and Ceftaroline. For the referenced antibiotics, the firm code in the reference table.
- Recalling firm
- Biomerieux Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-06-14
- Terminated
- 2015-06-01
- Firm location
- Hazelwood, MO, United States
Product description
bioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDT V2S 6.01 PT. Part number: 414566. The product is used with the VITEK 2 automated system which consists of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.
Reason for recall
A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bid-directional Computer interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin and Ceftaroline. For the referenced antibiotics, the firm code in the reference table
Data sourced from openFDA. This site is unofficial and independent of the FDA.