Recall Z-1912-2020— ICU Medical, Inc.

Class II · Terminated

Class II FDA medical device recall by ICU Medical, Inc., initiated 2019-10-04, terminated 2021-12-28. FDA's recall record names no specific 510(k) clearance for this event. Product: REF: B30211, 37" (94 cm) Appx 4.2 ml, 10 Drop Admin Set W/Luer Lock, Bag Hanger, Sterile R, UDI: (01)00887709063211. Reason: There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences..

Recalling firm
ICU Medical, Inc.
Classification
Class II
Status
Terminated
Initiated
2019-10-04
Terminated
2021-12-28
Firm location
San Clemente, CA, United States

Product description

REF: B30211, 37" (94 cm) Appx 4.2 ml, 10 Drop Admin Set W/Luer Lock, Bag Hanger, Sterile R, UDI: (01)00887709063211

Reason for recall

There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.