Recall Z-1906-2021— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2021-05-07, terminated 2023-03-27. FDA's recall record names no specific 510(k) clearance for this event. Product: Access SARS-CoV-2 lgG II Calibrator, REF C69058, IVD, CE UDI: (01)15099590742751. Reason: SARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2021-05-07
Terminated
2023-03-27
Firm location
Brea, CA, United States

Product description

Access SARS-CoV-2 lgG II Calibrator, REF C69058, IVD, CE UDI: (01)15099590742751

Reason for recall

SARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.

Data sourced from openFDA. This site is unofficial and independent of the FDA.