Recall Z-1905-2020— Medtronic Navigation, Inc.
Class II · Terminated
Class II FDA medical device recall by Medtronic Navigation, Inc., initiated 2019-07-08, terminated 2024-10-16. It names one FDA 510(k) clearance: K071328. Product: Visualase Cooled Laser Applicator System Tubing Set kits Product numbers: VCLAS 9735559, 3mm tip/1.65mm catheter VCLAS 9735560, 10mm tip/1.65mm catheter VCLAS 9735561, 15mm tip/1.85mm catheter. Reason: The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow..
- Recalling firm
- Medtronic Navigation, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-07-08
- Terminated
- 2024-10-16
- Firm location
- Louisville, CO, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Visualase Cooled Laser Applicator System Tubing Set kits Product numbers: VCLAS 9735559, 3mm tip/1.65mm catheter VCLAS 9735560, 10mm tip/1.65mm catheter VCLAS 9735561, 15mm tip/1.85mm catheter
Reason for recall
The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow.