Recall Z-1903-2019— Siemens Medical Solutions USA, Inc
Class II · Terminated
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2019-06-06, terminated 2021-02-09. It names one FDA 510(k) clearance: K181407. Product: Siemens Artis Q.zen ceiling-Interventional Fluoroscopic X-Ray System Model Number: 10848354. Reason: Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel.
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-06-06
- Terminated
- 2021-02-09
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Siemens Artis Q.zen ceiling-Interventional Fluoroscopic X-Ray System Model Number: 10848354
Reason for recall
Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel
Data sourced from openFDA. This site is unofficial and independent of the FDA.