Recall Z-1899-2024— Lucid Diagnostics, Inc.

Class II · Ongoing

Class II FDA medical device recall by Lucid Diagnostics, Inc., initiated 2024-04-24. It names one FDA 510(k) clearance: K203450. Product: Brand Name: EsophaCap Product Name: EsophaCap (20mm diameter, 20 pores/inch) Model/Catalog Number: EC20T20VI Software Version: N/A Product Description: EsophaCap is pouched as individual units. 10 units go into a shelf-box. Component: No. Reason: The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper esophageal sphincter between the pharynx and the esophagus, which is the narrowest part of the alimentary canal..

Recalling firm
Lucid Diagnostics, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-04-24
Firm location
Foxboro, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brand Name: EsophaCap Product Name: EsophaCap (20mm diameter, 20 pores/inch) Model/Catalog Number: EC20T20VI Software Version: N/A Product Description: EsophaCap is pouched as individual units. 10 units go into a shelf-box. Component: No

Reason for recall

The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper esophageal sphincter between the pharynx and the esophagus, which is the narrowest part of the alimentary canal.