Recall Z-1899-2017— Roche Diabetes Care, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diabetes Care, Inc., initiated 2017-03-14, terminated 2018-01-16. It names one FDA 510(k) clearance: K150910. Product: Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android Product Usage: This is a digital product available for download direct to customers from the Apple App Store and the Google Play Store. The total number of downloads is not available. The number of unique users for each affected version from the date of initial distribution through 03/23/2017 is noted in 7a. However, some users will have used all three affected versions as they upgraded from version to version when the versions became available and many others will have tried the app and discontinued use. In the last 30 days, 02/22/2017 03/23/2017, there have been 16,781 unique users in the US with 264 of those seeking bolus advice and 19,268 unique users in the rest of world (excluding Canada) with 415 of those users seeking bolus advice.. Reason: A program error (bug) in the Bolus Advisor feature, which could result in incorrect bolus advice and a potential insulin over-delivery..

Recalling firm
Roche Diabetes Care, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-03-14
Terminated
2018-01-16
Firm location
Indianapolis, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android Product Usage: This is a digital product available for download direct to customers from the Apple App Store and the Google Play Store. The total number of downloads is not available. The number of unique users for each affected version from the date of initial distribution through 03/23/2017 is noted in 7a. However, some users will have used all three affected versions as they upgraded from version to version when the versions became available and many others will have tried the app and discontinued use. In the last 30 days, 02/22/2017 03/23/2017, there have been 16,781 unique users in the US with 264 of those seeking bolus advice and 19,268 unique users in the rest of world (excluding Canada) with 415 of those users seeking bolus advice.

Reason for recall

A program error (bug) in the Bolus Advisor feature, which could result in incorrect bolus advice and a potential insulin over-delivery.