Recall Z-1898-2020— NuVasive Inc

Class II · Terminated

Class II FDA medical device recall by NuVasive Inc, initiated 2020-02-13, terminated 2022-05-25. It names one FDA 510(k) clearance: K171791. Product: NUVASIVE MAGEC 2 Rod X.X mm XX mm, REF MC2-XXXXS LOT 0000000 AAA - Product Usage: comprised of a sterile single-use spinal rod that is surgically implanted using appropriate NuVasive Reline, Reline Small Stature, and Armada fixation components (i.e. Pedicle screws, hooks and/or connectors, provided non-sterile, to be sterilized by the end user).. Reason: Post-implantation separation of an actuator end cap component that may expose internal components of the actuator, which could lead to hastened degeneration of the internal components and egress of Titanium alloy wear debris and resultant localized tissue discoloration..

Recalling firm
NuVasive Inc
Classification
Class II
Status
Terminated
Initiated
2020-02-13
Terminated
2022-05-25
Firm location
San Diego, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

NUVASIVE MAGEC 2 Rod X.X mm XX mm, REF MC2-XXXXS LOT 0000000 AAA - Product Usage: comprised of a sterile single-use spinal rod that is surgically implanted using appropriate NuVasive Reline, Reline Small Stature, and Armada fixation components (i.e. Pedicle screws, hooks and/or connectors, provided non-sterile, to be sterilized by the end user).

Reason for recall

Post-implantation separation of an actuator end cap component that may expose internal components of the actuator, which could lead to hastened degeneration of the internal components and egress of Titanium alloy wear debris and resultant localized tissue discoloration.

Data sourced from openFDA. This site is unofficial and independent of the FDA.