Recall Z-1896-2020— Intuitive Surgical Inc

Class II · Terminated

Class II FDA medical device recall by Intuitive Surgical Inc, initiated 2019-04-11, terminated 2024-04-08. It names one FDA 510(k) clearance: K131861. Product: da Vinci Xi Surgical System, IS4000 - Product Usage: The da Vinci Xi surgical system (IS4000) is made of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The Patient Cart is the operative component of the IS4000 system and its primary function is to support the instrument arms and camera arm. The system has four Universal Surgical Manipulators (USM) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. There are 4 USM per da Vinci Xi surgical system.. Reason: Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking..

Recalling firm
Intuitive Surgical Inc
Classification
Class II
Status
Terminated
Initiated
2019-04-11
Terminated
2024-04-08
Firm location
Sunnyvale, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

da Vinci Xi Surgical System, IS4000 - Product Usage: The da Vinci Xi surgical system (IS4000) is made of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The Patient Cart is the operative component of the IS4000 system and its primary function is to support the instrument arms and camera arm. The system has four Universal Surgical Manipulators (USM) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. There are 4 USM per da Vinci Xi surgical system.

Reason for recall

Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking.

Data sourced from openFDA. This site is unofficial and independent of the FDA.