Recall Z-1890-2017— Medtronic Inc.

Class II · Terminated

Class II FDA medical device recall by Medtronic Inc., initiated 2017-01-31, terminated 2017-05-04. FDA's recall record names no specific 510(k) clearance for this event. Product: MiniMed 620G and MiniMed 640G Insulin Pumps with Software Version 2.6 and below are affected. Catalog Numbers for these insulin pumps are MMT-1510, MMT-1710, MMT-1711, MMT-1712.. Reason: Medtronic has identified a software anomaly that can prevent the internal battery of the pump from charging. When the software anomaly occurs it leads to battery depletion and pump shutdown..

Recalling firm
Medtronic Inc.
Classification
Class II
Status
Terminated
Initiated
2017-01-31
Terminated
2017-05-04
Firm location
Northridge, CA, United States

Product description

MiniMed 620G and MiniMed 640G Insulin Pumps with Software Version 2.6 and below are affected. Catalog Numbers for these insulin pumps are MMT-1510, MMT-1710, MMT-1711, MMT-1712.

Reason for recall

Medtronic has identified a software anomaly that can prevent the internal battery of the pump from charging. When the software anomaly occurs it leads to battery depletion and pump shutdown.