Recall Z-1886-2017— Stryker Instruments Div. of Stryker Corporation

Class II · Terminated

Class II FDA medical device recall by Stryker Instruments Div. of Stryker Corporation, initiated 2017-02-28, terminated 2023-01-23. FDA's recall record names no specific 510(k) clearance for this event. Product: Stryker SmartLife Large Aseptic Housing (7126-120-000). Reason: Stryker Instruments issued a notice of correction for an update to cleaning practices for the Stryker SmartLife Large Aseptic Housings following complaints of deterioration to the housing and the top and bottom housing separating at the weld. Using cleaning practices outside of the HIFU could lead to housing separating at the weld..

Recalling firm
Stryker Instruments Div. of Stryker Corporation
Classification
Class II
Status
Terminated
Initiated
2017-02-28
Terminated
2023-01-23
Firm location
Portage, MI, United States

Product description

Stryker SmartLife Large Aseptic Housing (7126-120-000)

Reason for recall

Stryker Instruments issued a notice of correction for an update to cleaning practices for the Stryker SmartLife Large Aseptic Housings following complaints of deterioration to the housing and the top and bottom housing separating at the weld. Using cleaning practices outside of the HIFU could lead to housing separating at the weld.