Recall Z-1882-2018— Visaris DOO
Class II · Terminated
Class II FDA medical device recall by Visaris DOO, initiated 2018-04-05, terminated 2019-09-19. It names one FDA 510(k) clearance: K160620. Product: Visaris Vision/Auto, ddRAura OTC APS, Diagnostic X-Ray System, Version 2.4 The Vision systems are intended for use in all areas of diagnostic radiography imaging and specifically at secondary and tertiary levels of medical care. Powerful positioning capability makes this system particularly suited for institutions with immobile or hardly mobile patients, such as trauma wards and emergency departments. The system s versatility also makes it suitable for unified diagnostic imaging services for medium size hospitals where this system can cover all diagnostic radiography needs. In general, Vision systems are intended to position the patient, x-ray source and digital detector in such a way as to expose only the desired patient anatomy to X-rays and capture the resulting x-ray image. They are then intended to process the acquired image to a diagnostic viewing level and export it in DICOM format through an option of several distribution media including film, paper, CD-DVD and digital PACS archives. The system is intended for routine, daily, dedicated use in radiographic diagnosis. It can be used in any type of specialized radiology wards or for general radiographic diagnostics. Reason: A failure of the tube arm suspension is possible while the tube arm height is being adjusted. During tube height adjustment there is a small probability of failure of both the primary and redundant holding mechanisms, causing uncontrollable descent of the tube arm and a potential risk of injury to the patient..
- Recalling firm
- Visaris DOO
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-04-05
- Terminated
- 2019-09-19
- Firm location
- Belgrade
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Visaris Vision/Auto, ddRAura OTC APS, Diagnostic X-Ray System, Version 2.4 The Vision systems are intended for use in all areas of diagnostic radiography imaging and specifically at secondary and tertiary levels of medical care. Powerful positioning capability makes this system particularly suited for institutions with immobile or hardly mobile patients, such as trauma wards and emergency departments. The system s versatility also makes it suitable for unified diagnostic imaging services for medium size hospitals where this system can cover all diagnostic radiography needs. In general, Vision systems are intended to position the patient, x-ray source and digital detector in such a way as to expose only the desired patient anatomy to X-rays and capture the resulting x-ray image. They are then intended to process the acquired image to a diagnostic viewing level and export it in DICOM format through an option of several distribution media including film, paper, CD-DVD and digital PACS archives. The system is intended for routine, daily, dedicated use in radiographic diagnosis. It can be used in any type of specialized radiology wards or for general radiographic diagnostics
Reason for recall
A failure of the tube arm suspension is possible while the tube arm height is being adjusted. During tube height adjustment there is a small probability of failure of both the primary and redundant holding mechanisms, causing uncontrollable descent of the tube arm and a potential risk of injury to the patient.
Data sourced from openFDA. This site is unofficial and independent of the FDA.