Recall Z-1882-2016— Greatbatch Medical
Class II · Terminated
Class II FDA medical device recall by Greatbatch Medical, initiated 2016-04-26, terminated 2018-06-07. FDA's recall record names no specific 510(k) clearance for this event. Product: Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000, TLT911-50-S20 TLT911-50-S20, TLT911-15-S01 TLT911-15-S01, TLTS0007-20-S19 TLTS0007-20-S19, TLT911-35-S01 TLT911-35-S01.. Reason: Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device..
- Recalling firm
- Greatbatch Medical
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-04-26
- Terminated
- 2018-06-07
- Firm location
- Plymouth, MN, United States
Product description
Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000, TLT911-50-S20 TLT911-50-S20, TLT911-15-S01 TLT911-15-S01, TLTS0007-20-S19 TLTS0007-20-S19, TLT911-35-S01 TLT911-35-S01.
Reason for recall
Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.