Recall Z-1879-2016— Elekta, Inc.

Class II · Terminated

Class II FDA medical device recall by Elekta, Inc., initiated 2016-05-18, terminated 2021-09-15. It names one FDA 510(k) clearance: K151233. Product: Monaco RTP System Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.. Reason: When DICOM exporting a 3D Monaco plan and the "Composite Field Sequencing" (CFS) checkbox is selected, the Y jaws will snap to the port shape when they should remain where they were defined..

Recalling firm
Elekta, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-05-18
Terminated
2021-09-15
Firm location
Atlanta, GA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Monaco RTP System Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Reason for recall

When DICOM exporting a 3D Monaco plan and the "Composite Field Sequencing" (CFS) checkbox is selected, the Y jaws will snap to the port shape when they should remain where they were defined.