Recall Z-1878-2019— GE Healthcare, LLC

Class III · Terminated

Class III FDA medical device recall by GE Healthcare, LLC, initiated 2019-05-10, terminated 2024-05-24. FDA's recall record names no specific 510(k) clearance for this event. Product: Leadwire sets with snap ends: REF/Catalog Number Description 2106381-001 ECG Leadwire set, 5-lead, snap, AHA, 74 cm/ 29 in 2106381-002 ECG Leadwire set, 5-lead, snap, AHA, 130 cm/ 51 in 2106381-003 ECG Leadwire set, 5-lead, snap, IEC, 74 cm/ 29 in 2106381-004 ECG Leadwire set, 5-lead, snap, IEC, 130 cm/ 51 in 2106381-005 ECG Leadwire set, 5-lead, snap, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106383-001 ECG Leadwire set, 5-lead, grouped, snap, AHA, 74 cm/ 29 in 2106383-002 ECG Leadwire set, 5-lead, grouped, snap, AHA, 130 cm/ 51 in 2106383-003 ECG Leadwire set, 5-lead, grouped, snap, IEC, 74 cm/ 29 in 2106383-004 ECG Leadwire set, 5-lead, grouped, snap, IEC, 130 cm/ 51 in 2106383-005 ECG Leadwire set, 5-lead, grouped, snap, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106383-006 ECG Leadwire set, 5-lead, grouped, snap, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106385-001 ECG Leadwire set, 3-lead, snap, AHA, 74 cm/ 29 in 2106385-002 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, AHA, 130 CM/ 51 IN 2106385-003 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, IEC, 74 CM/ 29 IN 2106385-004 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, IEC, 130 CM/ 51 IN 2106394-003 ECG Leadwire set, 5-lead V2-6, snap, AHA, 74 cm/ 29 in 2106395-001 ECG Leadwire set, 6-lead, grouped, snap, AHA, 74 cm/ 29 in 2106396-001 ECG Leadwire set, 6-lead, grouped, snap, IEC, 74 cm/ 29 in 2106396-002 ECG Leadwire set, 6-lead, grouped, snap, IEC, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.. Reason: ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm..

Recalling firm
GE Healthcare, LLC
Classification
Class III
Status
Terminated
Initiated
2019-05-10
Terminated
2024-05-24
Firm location
Waukesha, WI, United States

Product description

Leadwire sets with snap ends: REF/Catalog Number Description 2106381-001 ECG Leadwire set, 5-lead, snap, AHA, 74 cm/ 29 in 2106381-002 ECG Leadwire set, 5-lead, snap, AHA, 130 cm/ 51 in 2106381-003 ECG Leadwire set, 5-lead, snap, IEC, 74 cm/ 29 in 2106381-004 ECG Leadwire set, 5-lead, snap, IEC, 130 cm/ 51 in 2106381-005 ECG Leadwire set, 5-lead, snap, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106383-001 ECG Leadwire set, 5-lead, grouped, snap, AHA, 74 cm/ 29 in 2106383-002 ECG Leadwire set, 5-lead, grouped, snap, AHA, 130 cm/ 51 in 2106383-003 ECG Leadwire set, 5-lead, grouped, snap, IEC, 74 cm/ 29 in 2106383-004 ECG Leadwire set, 5-lead, grouped, snap, IEC, 130 cm/ 51 in 2106383-005 ECG Leadwire set, 5-lead, grouped, snap, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106383-006 ECG Leadwire set, 5-lead, grouped, snap, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106385-001 ECG Leadwire set, 3-lead, snap, AHA, 74 cm/ 29 in 2106385-002 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, AHA, 130 CM/ 51 IN 2106385-003 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, IEC, 74 CM/ 29 IN 2106385-004 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, IEC, 130 CM/ 51 IN 2106394-003 ECG Leadwire set, 5-lead V2-6, snap, AHA, 74 cm/ 29 in 2106395-001 ECG Leadwire set, 6-lead, grouped, snap, AHA, 74 cm/ 29 in 2106396-001 ECG Leadwire set, 6-lead, grouped, snap, IEC, 74 cm/ 29 in 2106396-002 ECG Leadwire set, 6-lead, grouped, snap, IEC, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

Reason for recall

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

Data sourced from openFDA. This site is unofficial and independent of the FDA.