Recall Z-1878-2013— Stryker Instruments Div. of Stryker Corporation
Class II · Terminated
Class II FDA medical device recall by Stryker Instruments Div. of Stryker Corporation, initiated 2013-06-24, terminated 2014-01-30. It names one FDA 510(k) clearance: K010309. Product: Sonopet Ultrasonic Aspirator Console (Console 110V with Pedal & Pole) Product Usage: The Sonopet Console is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft and hard tissue is desirable, including neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, plastic and reconstructive surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and thoracoscopic surgery.. Reason: A supplied component exhibiting corrosion was built into the power supply boards of the Sonopet Consoles listed above. The issue is limited to 2 lots of the supplied component. The presence of the corrosion on the relay of the power supply board can potentially lead to a conduction defect which could cause the irrigation pump to stop working. The purpose of the irrigation pump is to supply sali.
- Recalling firm
- Stryker Instruments Div. of Stryker Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-06-24
- Terminated
- 2014-01-30
- Firm location
- Portage, MI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Sonopet Ultrasonic Aspirator Console (Console 110V with Pedal & Pole) Product Usage: The Sonopet Console is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft and hard tissue is desirable, including neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, plastic and reconstructive surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and thoracoscopic surgery.
Reason for recall
A supplied component exhibiting corrosion was built into the power supply boards of the Sonopet Consoles listed above. The issue is limited to 2 lots of the supplied component. The presence of the corrosion on the relay of the power supply board can potentially lead to a conduction defect which could cause the irrigation pump to stop working. The purpose of the irrigation pump is to supply sali
Data sourced from openFDA. This site is unofficial and independent of the FDA.