Recall Z-1871-2024— Philips North America Llc
Class II · Ongoing
Class II FDA medical device recall by Philips North America Llc, initiated 2024-04-15. It names one FDA 510(k) clearance: K180015. Product: Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728143. Reason: Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity.
- Recalling firm
- Philips North America Llc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-04-15
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728143
Reason for recall
Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity