Recall Z-1869-2021— Covidien, LP
Class II · Ongoing
Class II FDA medical device recall by Covidien, LP, initiated 2021-04-29. It names one FDA 510(k) clearance: K941381. Product: HME for Tracheostomized Patients, Item Code 353S13046 - Product Usage: indicated for single use on anesthetized patients and respiratory care patients who require a breathing circuit with ISO standard 15mm or 22 mm connectors.. Reason: The company was notified by their supplier of sterilization services for airway products in Italy of potential deviations from validated parameters for ethylene oxide sterilization..
- Recalling firm
- Covidien, LP
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-04-29
- Firm location
- North Haven, CT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
HME for Tracheostomized Patients, Item Code 353S13046 - Product Usage: indicated for single use on anesthetized patients and respiratory care patients who require a breathing circuit with ISO standard 15mm or 22 mm connectors.
Reason for recall
The company was notified by their supplier of sterilization services for airway products in Italy of potential deviations from validated parameters for ethylene oxide sterilization.