Recall Z-1865-2026— Angiodynamics, Inc.

Class II · Ongoing

Class II FDA medical device recall by Angiodynamics, Inc., initiated 2026-03-03. It names 3 FDA 510(k) clearances: K173762, K161596, K112452. Product: Soft-Vu Angiographic Catheter, Cobra (2), Non-Braided; SOFT-VU CB2 4F X 100CM 038 NB 0SH; Catalog No.: 10714018; Product/UPN No.: H787107140185 (Box), H787107140180 (Pouch); Box Quantity: 5 pouches;. Reason: The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub..

Recalling firm
Angiodynamics, Inc.
Classification
Class II
Status
Ongoing
Initiated
2026-03-03
Firm location
Queensbury, NY, United States

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Soft-Vu Angiographic Catheter, Cobra (2), Non-Braided; SOFT-VU CB2 4F X 100CM 038 NB 0SH; Catalog No.: 10714018; Product/UPN No.: H787107140185 (Box), H787107140180 (Pouch); Box Quantity: 5 pouches;

Reason for recall

The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.

Data sourced from openFDA. This site is unofficial and independent of the FDA.