Recall Z-1863-2024— Philips North America Llc

Class II · Ongoing

Class II FDA medical device recall by Philips North America Llc, initiated 2024-05-03. FDA's recall record names no specific 510(k) clearance for this event. Product: SmartPath to Ingenia Elition X - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 782118. Reason: The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire.

Recalling firm
Philips North America Llc
Classification
Class II
Status
Ongoing
Initiated
2024-05-03
Firm location
Cambridge, MA, United States

Product description

SmartPath to Ingenia Elition X - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 782118

Reason for recall

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Data sourced from openFDA. This site is unofficial and independent of the FDA.