Recall Z-1863-2021— Vero Biotech, LLC

Class I · Terminated

Class I FDA medical device recall by Vero Biotech, LLC, initiated 2021-05-19, terminated 2023-10-12. FDA's recall record names no specific 510(k) clearance for this event. Product: GENOSYL DS; Nitric Oxide Delivery System Combination Product NDA202860. Reason: Reports received of NO measured below desired dose during transition between primary console and backup console since Software Version 2.2.3 was uploaded to consoles in the field..

Recalling firm
Vero Biotech, LLC
Classification
Class I
Status
Terminated
Initiated
2021-05-19
Terminated
2023-10-12
Firm location
Atlanta, GA, United States

Product description

GENOSYL DS; Nitric Oxide Delivery System Combination Product NDA202860

Reason for recall

Reports received of NO measured below desired dose during transition between primary console and backup console since Software Version 2.2.3 was uploaded to consoles in the field.