Recall Z-1863-2014— Philips Medical Systems, Inc.

Class II · Terminated

Class II FDA medical device recall by Philips Medical Systems, Inc., initiated 2014-06-13, terminated 2016-01-27. It names one FDA 510(k) clearance: K110151. Product: Ingenia 1.5T R5, magnetic resonance imaging system.. Reason: For some identified Ingenia systems shipped between December 2013 and March 2014, cover set rear magnetic M10 nuts were erroneously supplied instead of A4 stainless steel nuts..

Recalling firm
Philips Medical Systems, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-06-13
Terminated
2016-01-27
Firm location
Andover, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Ingenia 1.5T R5, magnetic resonance imaging system.

Reason for recall

For some identified Ingenia systems shipped between December 2013 and March 2014, cover set rear magnetic M10 nuts were erroneously supplied instead of A4 stainless steel nuts.

Data sourced from openFDA. This site is unofficial and independent of the FDA.