Recall Z-1862-2021— Circulatory Technology Inc.
Class II · Ongoing
Class II FDA medical device recall by Circulatory Technology Inc., initiated 2021-04-27. It names one FDA 510(k) clearance: K981284. Product: The Better-Bladder (BB), part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits. Model: BB14. Reason: The device is used as part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits and may collapse during use which can increase resistance to flow in the venous line and cause a drop in circuit blood flow..
- Recalling firm
- Circulatory Technology Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-04-27
- Firm location
- Oyster Bay, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
The Better-Bladder (BB), part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits. Model: BB14
Reason for recall
The device is used as part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits and may collapse during use which can increase resistance to flow in the venous line and cause a drop in circuit blood flow.