Recall Z-1862-2012— Stryker Medical Division of Stryker Corporation
Class II · Terminated
Class II FDA medical device recall by Stryker Medical Division of Stryker Corporation, initiated 2012-05-17, terminated 2014-05-12. FDA's recall record names no specific 510(k) clearance for this event. Product: Stryker MA204 Bed Model (FL25E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone. Reason: An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac.
- Recalling firm
- Stryker Medical Division of Stryker Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-05-17
- Terminated
- 2014-05-12
- Firm location
- Portage, MI, United States
Product description
Stryker MA204 Bed Model (FL25E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone
Reason for recall
An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac
Data sourced from openFDA. This site is unofficial and independent of the FDA.