Recall Z-1860-2024— Philips North America Llc
Class II · Ongoing
Class II FDA medical device recall by Philips North America Llc, initiated 2024-05-03. FDA's recall record names no specific 510(k) clearance for this event. Product: SmartPath to dStream for 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 782146; 782112; 781260. Reason: The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire.
- Recalling firm
- Philips North America Llc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-05-03
- Firm location
- Cambridge, MA, United States
Product description
SmartPath to dStream for 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 782146; 782112; 781260
Reason for recall
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Data sourced from openFDA. This site is unofficial and independent of the FDA.