Recall Z-1859-2014— NxStage Medical, Inc.

Class II · Terminated

Class II FDA medical device recall by NxStage Medical, Inc., initiated 2014-05-14, terminated 2017-03-13. It names 2 FDA 510(k) clearances: K140571, K111174. Product: NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-306 Lactate 45, Potassium 2, Sodium 140, Calcium 3, Magnesium 1, Chloride 101, Glucose (mg/dL) 100 Batch Size (liters) 50 The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch.. Reason: SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification.

Recalling firm
NxStage Medical, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-05-14
Terminated
2017-03-13
Firm location
Lawrence, MA, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-306 Lactate 45, Potassium 2, Sodium 140, Calcium 3, Magnesium 1, Chloride 101, Glucose (mg/dL) 100 Batch Size (liters) 50 The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch.

Reason for recall

SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification