Recall Z-1858-2022— Hologic, Inc
Class II · Ongoing
Class II FDA medical device recall by Hologic, Inc, initiated 2022-08-09. It names one FDA 510(k) clearance: K152070. Product: The BioZorb LP Marker is an implantable radiopaque marker used to facilitate visualization of a soft tissue site. The BioZorb LP Marker is comprised of a bioabsorbable PLA (polylactic acid) component which resorbs completely in 1 year or more and a permanent component (titanium). The BioZorb LP Marker is provided sterile for single use and is implantable.. Reason: Hologic BioZorb LP Marker, (Product number F0221), lot number 22A0RL, contained removable labels with an expiration date that indicated the product had already expired. The expiration date on the removable label is actually manufacturing date, and it's different from the one on the shelf box label, which is the correct expiration date..
- Recalling firm
- Hologic, Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-08-09
- Firm location
- Marlborough, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
The BioZorb LP Marker is an implantable radiopaque marker used to facilitate visualization of a soft tissue site. The BioZorb LP Marker is comprised of a bioabsorbable PLA (polylactic acid) component which resorbs completely in 1 year or more and a permanent component (titanium). The BioZorb LP Marker is provided sterile for single use and is implantable.
Reason for recall
Hologic BioZorb LP Marker, (Product number F0221), lot number 22A0RL, contained removable labels with an expiration date that indicated the product had already expired. The expiration date on the removable label is actually manufacturing date, and it's different from the one on the shelf box label, which is the correct expiration date.