Recall Z-1857-2022— MEDLINE INDUSTRIES, LP - Northfield
Class II · Ongoing
Class II FDA medical device recall by MEDLINE INDUSTRIES, LP - Northfield, initiated 2022-08-03. FDA's recall record names no specific 510(k) clearance for this event. Product: MEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14; b. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14T. Reason: Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process..
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-08-03
- Firm location
- Northfield, IL, United States
Product description
MEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14; b. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14T
Reason for recall
Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.