Recall Z-1851-2024— Philips North America Llc
Class II · Ongoing
Class II FDA medical device recall by Philips North America Llc, initiated 2024-05-03. It names 2 FDA 510(k) clearances: K213583, K193215. Product: Ingenia Ambition X- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 782138; (2) 782109; (3) 781356. Reason: The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire.
- Recalling firm
- Philips North America Llc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-05-03
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
2 on FDA's record- K213583 — Achieva, Ingenia, Ingenia CX, Ingenia Elition and Ingenia Ambition MR Systems
- K193215 — Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Ingenia Ambition X- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 782138; (2) 782109; (3) 781356
Reason for recall
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Data sourced from openFDA. This site is unofficial and independent of the FDA.