Recall Z-1848-2014— SteriGear, LLC

Class II · Terminated

Class II FDA medical device recall by SteriGear, LLC, initiated 2011-01-11, terminated 2015-01-02. FDA's recall record names no specific 510(k) clearance for this event. Product: SteriGear Urinary Drain Bag with Fig Leaf Cover, 2000 ml, REF 10270, The Fig Leaf, SteriGear 362 S. University Avenue, provo, UT 84601, 1-800-398-3259, SteriGear.com, Sterile, EO Urinary Drainage Collection Kit, for Indwelling Catheter. This device is intended to collect the Urine drained from an indwelling catheter. Reason: SteriGear notifed customers of circumstances in which The Fig Leaf urinary drain bag valve can close creating a negative pressure that can suck the non-return valve closed..

Recalling firm
SteriGear, LLC
Classification
Class II
Status
Terminated
Initiated
2011-01-11
Terminated
2015-01-02
Firm location
Provo, UT, United States

Product description

SteriGear Urinary Drain Bag with Fig Leaf Cover, 2000 ml, REF 10270, The Fig Leaf, SteriGear 362 S. University Avenue, provo, UT 84601, 1-800-398-3259, SteriGear.com, Sterile, EO Urinary Drainage Collection Kit, for Indwelling Catheter. This device is intended to collect the Urine drained from an indwelling catheter

Reason for recall

SteriGear notifed customers of circumstances in which The Fig Leaf urinary drain bag valve can close creating a negative pressure that can suck the non-return valve closed.

Data sourced from openFDA. This site is unofficial and independent of the FDA.