Recall Z-1846-2024— Philips North America Llc

Class II · Ongoing

Class II FDA medical device recall by Philips North America Llc, initiated 2024-05-03. It names one FDA 510(k) clearance: K193215. Product: Ingenia 1.5T CX - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 781262 (2) 781261. Reason: The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire.

Recalling firm
Philips North America Llc
Classification
Class II
Status
Ongoing
Initiated
2024-05-03
Firm location
Cambridge, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Ingenia 1.5T CX - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 781262 (2) 781261

Reason for recall

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Data sourced from openFDA. This site is unofficial and independent of the FDA.