Recall Z-1840-2019— CooperSurgical, Inc.
Class II · Terminated
Class II FDA medical device recall by CooperSurgical, Inc., initiated 2019-04-23, terminated 2020-04-24. It names one FDA 510(k) clearance: K963348. Product: LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components - LEEP PRECISION Cart, LEEP PRECISION Generator, and the LEEP PRECISION Smoke Evacuator for loop electrosurgical excision procedures. The loop electrosurgical excision procedure (LEEP) utilizes a very thin wire in the shape of a loop heated by electrical current generated by a power supply to efficiently remove abnormal cells. The LEEP PRECISION INTEGRATED SYSTEM has been designed to provide an efficient workstation that can efficiently meet all the unique LEEP requirements of the clinic and practice-based facilities .. Reason: 2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses..
- Recalling firm
- CooperSurgical, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-04-23
- Terminated
- 2020-04-24
- Firm location
- Trumbull, CT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components - LEEP PRECISION Cart, LEEP PRECISION Generator, and the LEEP PRECISION Smoke Evacuator for loop electrosurgical excision procedures. The loop electrosurgical excision procedure (LEEP) utilizes a very thin wire in the shape of a loop heated by electrical current generated by a power supply to efficiently remove abnormal cells. The LEEP PRECISION INTEGRATED SYSTEM has been designed to provide an efficient workstation that can efficiently meet all the unique LEEP requirements of the clinic and practice-based facilities .
Reason for recall
2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.