Recall Z-1822-2013— Steritec Products Mfc. Co., Inc.
Class II · Terminated
Class II FDA medical device recall by Steritec Products Mfc. Co., Inc., initiated 2013-06-10, terminated 2013-12-19. It names one FDA 510(k) clearance: K930682. Product: Medline Steam Biological Indicators, SteriTec Product Code PL 353, Distributor Product Code MDS200525 EZTest Steam Biological Indicator: Mesa EZTest Steam Biological Indicator is a self contained Biological Indicator intended for monitoring the efficacy of 121¿C, 132¿C, 134¿C and 135¿C steam sterilization cycles. Mesa EZTest has a recommended incubation time of 24 hours.. Reason: Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims..
- Recalling firm
- Steritec Products Mfc. Co., Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-06-10
- Terminated
- 2013-12-19
- Firm location
- Englewood, CO, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Medline Steam Biological Indicators, SteriTec Product Code PL 353, Distributor Product Code MDS200525 EZTest Steam Biological Indicator: Mesa EZTest Steam Biological Indicator is a self contained Biological Indicator intended for monitoring the efficacy of 121¿C, 132¿C, 134¿C and 135¿C steam sterilization cycles. Mesa EZTest has a recommended incubation time of 24 hours.
Reason for recall
Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.