Recall Z-1821-2012— Philips Healthcare Inc.
Class II · Terminated
Class II FDA medical device recall by Philips Healthcare Inc., initiated 2012-05-11, terminated 2016-10-06. It names 2 FDA 510(k) clearances: K031535, K970640. Product: Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system. Reason: If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut.
- Recalling firm
- Philips Healthcare Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-05-11
- Terminated
- 2016-10-06
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system
Reason for recall
If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut