Recall Z-1818-2025— Onkos Surgical, Inc.
Class II · Ongoing
Class II FDA medical device recall by Onkos Surgical, Inc., initiated 2023-12-22. It names one FDA 510(k) clearance: K203815. Product: ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage System with BioGrip.. Reason: Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems..
- Recalling firm
- Onkos Surgical, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-12-22
- Firm location
- Parsippany, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage System with BioGrip.
Reason for recall
Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems.