Recall Z-1818-2015— Hill-Rom, Inc.

Class II · Terminated

Class II FDA medical device recall by Hill-Rom, Inc., initiated 2015-03-20, terminated 2017-08-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Viking M, L, and XL Mobile Lifts. Non-AC-powered patient lift.. Reason: Complaints (including one reported death) allegedly of the lift arm drifting down suddenly. If the lift arm assembly is manually lifted, the actuator can become damaged and get stuck in the highest position. If a patient is lifted into the sling while the actuator is stuck, there is a potential risk of a free fall of the patient, resulting in minor or potentially catastrophic injuries..

Recalling firm
Hill-Rom, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-03-20
Terminated
2017-08-04
Firm location
Batesville, IN, United States

Product description

Viking M, L, and XL Mobile Lifts. Non-AC-powered patient lift.

Reason for recall

Complaints (including one reported death) allegedly of the lift arm drifting down suddenly. If the lift arm assembly is manually lifted, the actuator can become damaged and get stuck in the highest position. If a patient is lifted into the sling while the actuator is stuck, there is a potential risk of a free fall of the patient, resulting in minor or potentially catastrophic injuries.

Data sourced from openFDA. This site is unofficial and independent of the FDA.