Recall Z-1818-2012— Philips Healthcare Inc.

Class II · Terminated

Class II FDA medical device recall by Philips Healthcare Inc., initiated 2012-05-22, terminated 2017-02-24. It names one FDA 510(k) clearance: K982795. Product: Philips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon Stationary Fluoroscopic system. Reason: With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "right patient side" marker, although this can appear on the left patient side.

Recalling firm
Philips Healthcare Inc.
Classification
Class II
Status
Terminated
Initiated
2012-05-22
Terminated
2017-02-24
Firm location
Andover, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Philips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon Stationary Fluoroscopic system

Reason for recall

With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "right patient side" marker, although this can appear on the left patient side