Recall Z-1817-2015— Boston Scientific Corporation

Class II · Terminated

Class II FDA medical device recall by Boston Scientific Corporation, initiated 2015-06-02, terminated 2016-10-26. It names one FDA 510(k) clearance: K150235. Product: CLEARSIGN II Amplifier for LABSYSTEM PRO EP Recording System, 120 channels, using firmware version 2.08. The channel numbers in the amplifier are as follows: Material number: H30120020210 - 40 Channel CLEARSIGN II Amplifier, catalog number 2002021; Material number: H30120020220 80 Channel CLEARSIGN II Amplifier, catalog number 2002022; Material number: H30120020230 120 Channel CLEARSIGN II Amplifier, catalog number 2002023; Material number: H30120020240 160 Channel Clearsign II Amplifier, catalog number 2002014. The CLEARSIGN II Amplifier is intended to amplify and condition electrocardiographic signals of biologic origin and pressure transducer input, transmitting this information to a host computer (the LABSYSTEM PRO EP Recording System) that can record and display the information. Reason: Boston Scientific has received complaints indicating that, when using the blood pressure (BP) channels on the CLEARSIGN II Amplifier, the surface Electrocardiogram (ECG) channels become over-written to a variable degree, with the result that it appears shifted from baseline on the system's output screen. This may, in turn, manifest as an uninterpretable ECG signal in the affected channel..

Recalling firm
Boston Scientific Corporation
Classification
Class II
Status
Terminated
Initiated
2015-06-02
Terminated
2016-10-26
Firm location
Lowell, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

CLEARSIGN II Amplifier for LABSYSTEM PRO EP Recording System, 120 channels, using firmware version 2.08. The channel numbers in the amplifier are as follows: Material number: H30120020210 - 40 Channel CLEARSIGN II Amplifier, catalog number 2002021; Material number: H30120020220 80 Channel CLEARSIGN II Amplifier, catalog number 2002022; Material number: H30120020230 120 Channel CLEARSIGN II Amplifier, catalog number 2002023; Material number: H30120020240 160 Channel Clearsign II Amplifier, catalog number 2002014. The CLEARSIGN II Amplifier is intended to amplify and condition electrocardiographic signals of biologic origin and pressure transducer input, transmitting this information to a host computer (the LABSYSTEM PRO EP Recording System) that can record and display the information

Reason for recall

Boston Scientific has received complaints indicating that, when using the blood pressure (BP) channels on the CLEARSIGN II Amplifier, the surface Electrocardiogram (ECG) channels become over-written to a variable degree, with the result that it appears shifted from baseline on the system's output screen. This may, in turn, manifest as an uninterpretable ECG signal in the affected channel.