Recall Z-1815-2017— RAYSEARCH LABORATORIES AB
Class II · Terminated
Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2017-01-18, terminated 2019-06-12. It names one FDA 510(k) clearance: K160093. Product: Radiation Therapy Treatment Planning System, Model 4.5, 4.7, 5.0 RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. Reason: An error in NVIDIA GPU (Graphics Processing Unit) card drivers can occur for certain software programs. In a non-standard Citrix environment where the GPU settings are configured with a virtual Citrix graphics board (graphics card) an erroneous calculation may be allowed..
- Recalling firm
- RAYSEARCH LABORATORIES AB
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-01-18
- Terminated
- 2019-06-12
- Firm location
- Stockholm, N/A, Sweden
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Radiation Therapy Treatment Planning System, Model 4.5, 4.7, 5.0 RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs
Reason for recall
An error in NVIDIA GPU (Graphics Processing Unit) card drivers can occur for certain software programs. In a non-standard Citrix environment where the GPU settings are configured with a virtual Citrix graphics board (graphics card) an erroneous calculation may be allowed.