Recall Z-1812-2020— WELCH ALLYN, INC/MORTARA
Class II · Terminated
Class II FDA medical device recall by WELCH ALLYN, INC/MORTARA, initiated 2020-03-18, terminated 2026-02-09. It names one FDA 510(k) clearance: K050896. Product: H12+ and patient cable for the H12+ Holter Recorders with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP. The 24-hour H12+ recorder uses a single AA alkaline battery to provide continuous 12-lead data recorded over a 24-hour period and a removable 24-hour compact flash (CF) card for data storage.. Reason: Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment..
- Recalling firm
- WELCH ALLYN, INC/MORTARA
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-03-18
- Terminated
- 2026-02-09
- Firm location
- Milwaukee, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
H12+ and patient cable for the H12+ Holter Recorders with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP. The 24-hour H12+ recorder uses a single AA alkaline battery to provide continuous 12-lead data recorded over a 24-hour period and a removable 24-hour compact flash (CF) card for data storage.
Reason for recall
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.